| Customization: | Available |
|---|---|
| Classification: | Biochemical Apparatus |
| Type: | Biochemistry Analyzer |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
| Index | Parameter | |
Basic info |
Wavelength Range | 12 Wavelength can be tested simultaneously by the concave-holographic image flattening-grating optical system |
| Reaction Temp | 37ºC. Solid direct-heating method | |
| Temp Accuracy | ±0.1ºC. Precise control with the PID technology. | |
| Testing Projects | Up to 80 colorimetry projects, and 3 ion testing projects | |
| Testing Method | One-point/two-point terminal method, two-point rate method, rate A method | |
| Testing Speed | 200 regular tests per hour, and 300 ion tests per hour (Na/K/Cl) | |
Sample &Reagent System |
Sample&Reagent Plate | Using the same plate for both sample and reagent |
| Reagent Site | 80 | |
| Reagent Volume | 10~300μL, 0.5μL per step | |
| Reagent Bottle volume | 20mL, 35mL, 70mL | |
| Storage Temp | 2~12ºC, also be customizable | |
| Sample Site | 40 All sites can be set as regular, emergency, calibration and QC | |
| Sample Container | Suitable for φ13 sample tube and 2mL standard tube | |
| Sample Type | Serum, whole blood, urine, gastric juice, cerebrospinal fluid, hydrothorax and ascite | |
| Sample Volume | 2~30μL, 0.1μL per step | |
| Liquid Level Sensor | Digital sensor with anti-collision function | |
| Barcode Recognition System | Optional equipping built-in barcode recognition module | |
| Data System |
Interface | RJ45 Ethernet interface |
| Computer | Optional equipping the computer and monitor | |
| LIS System | Available for LIS/HIS connection | |
Diagnostic System |
Cuvette Type | Quartz |
| Cuvette Light Path | 5mm | |
| Reaction Cuvette Sites | 56 | |
| Reaction Volume | 100~360μL | |
| Light Source | 20W/12V Long lived quartz halogen tungsten lamp with air cooling system | |
| Absorbance linear range | 0~3.6Abs | |
| ontamination Rate | <0.1% | |
| Quality Control | Manual and automatic processing the linear and nonlinear calibration; two-concentration method for the real-time QC, intra-day QC, cross-day | |
| Auto Flushing | Automatically washing the reaction cuvette | |
| Mixture System | Separately mixture after adding the reagent | |
| Instrument Parameters |
Weight | ≈100Kg |
| Dimension | 820*716*506mm (L×W×H) | |
| TDP | 500VA | |